Energy-Dispersive X-ray Fluorescence Spectrometer MARS
TECHNICAL AND METROLOGICAL DETAILS
Analyzed elements | from Na to U (air) from B to U (vacuum) |
Detection limit (depending on the elements), % | 0,0001 ÷ 100 (solid) |
Types of samples | solid, powders, liquid, films |
Measurement environment | air, vacuum(option) |
Detector Type | SDD |
Cooling | Peltier elements |
Energy Resolution @ 5.9 keV | 125 ÷ 135 eV |
X-ray tube combined with power source (monoblock) | Yes |
Target Material | Мo |
Option | Rh, Ag, Pd, Au, W |
Applying filters | Yes |
Tube voltage | 10 to 40 (60) kV |
Tube current | 5 to 200 µA |
Cooling | Air Cooled |
Connectivity | USB |
Spectrometer Control | Software with automatic shutdown |
Input supply | 110 to 250 V AC / 50-60 Hz, battery option |
Sample sizes, mm | 150 x 150 mm |
Weight, kg | 4 |
Dimensions (DxWxH), mm | 273х223х205 |